Изберете език

Aerosol Generator ATM 226

Производител: Topas
Product sheet ATM 200 series

ATM 226 Atomizer Aerosol Generator for aerosol generation from liquids for mobile use (e.g. cleanroom acceptance measurement), integrated compressor

Standards

VDI 4258-1
VDI 3491-2
DIN 12980
DIN EN 12469
ISO 14644-3
ISO 21083-1

The ATM 226 Atomizer Aerosol Generator is especially suited for clean room and filter testing applications according to international standards.

&Function test at laminar flow boxes (DIN EN 12469) and cytostatica safety workbenches (DIN 12980), the qualitative test of filters for suspended matter (HEPA and ULPA filters), and acceptance measurements in clean rooms (according to DIN ISO 14644 and VDI 2083) are a essential necessity for personal security and product safety.

Such measurements and checkouts need a suitable test aerosol. Therefore VDI-directive 3491 proposes an aerosol generator for producing a test aerosol with defined characteristics, like the ATM 226. By high generated particle numbers at a particle size distribution orientated at the Most Penetration Particle Size (MPPS) of the filter the locating of leaks in the system is possible fast and safe. If an oily aerosol substance is used (e.g. DOP, DEHS) the aerosol generator produces spherical particles mainly in the size range 0,1 … 0,5 µm with a calculable residence time. That does nearly not extra contaminate the filter material.

A further range of application for aerosol generators / liquid aerosol nebulizers is calibration of measuring devices and visualization of currents. That enables measurements of local velocities and flow profiles of the tracer particles by means of optical methods.

Benefits

  • compact stainless steel housing – suitable for clean room application
  • connection directly to mains supply
  • polydisperse aerosol, mainly < 1 µm
  • very high permanence of the generated particle size distribution
  • defined and high particle number concentration
  • easy handling, designed for mobile use (internal pump)
  • connection directly to mains voltage 100-240 VAC
  • increased mass output

Applications

  • testing of HEPA and ULPA filter media
  • acceptance measurements of clean rooms and safety cabinets
  • generation of tracer particles at low flow rates

Accessories

  • aerosol substance DEHS
  • diffusion dryer (DDU 570/L, DDU 570/H)
  • aerosol dilution system (DIL 554), stainless steel housing, battery operation
  • rectangular sampling probe (SYS 529) according to EN ISO 14644-3 and VDI 2083-3
  • cleanroom particle counter (LAP 340)
  • software for clean room validation (CRQWinⓇ)
  • arrying case
Parameter titleUnitValue
setting parameter - air flow rate
setting range Skt 5 ... 40 (scale mark on rotameter)
setting resolution Skt continously adjustable
setting method - manuell
aerosol, volumetric flow rate L/h 70 ... 300
aerosol, mass flow rate (aerosol substance) g/h 0,5 ... 2,5
operating medium, gas/air - ambient air
operating medium, aerosol substance - liquids (DEHS, PAO, ...), solutions (NaCl, KCl, ...), suspensions (PSL, ...)
operating medium, aerosol substance - filling volume mL 20 ... 80
power supply - 100 ... 240 AC
hose connector mm Ø 8
power consumption* W max. 36
environmental condition, max. counterpressure kPa 20
dimensions (w × h × d) mm 120 × 195 × 300
weight kg 4,8