Aerosol Generator ATM 228

Производител: Topas
product sheet ATM 228
application sheet PTI/PEMS
application sheet CRE

The aerosol generator ATM 228 is a further development of the ATM 226 aerosol generator and serves for the mobile generation of test and calibration aerosols from pure liquids, solutions and suspensions. The generator complies to all requirements of VDI 3491-2.

Standards

VDI 4258-1
VDI 3491-2
DIN 12980
DIN EN 12469
ISO 14644-3
ISO 21083-1
DIN EN 61577-4

Aerosol generation is realised by a brushless compressor. Accordingly, no external compressed air supply is required. For the purpose of reproducible and long-term stable aerosol generation, the air flow rate is controlled via the pressure drop over the nozzle. This allows also a reproducible and stable generator operation even at very low air flow rates and thus also at lowest particle production rates.

Benefits

  • stable aerosol generation even at lowest particle production rates
  • safe and reproducible adjustment of operating points
  • suitable for mobile operation (internal compressor, optional battery operation mode)
  • remote control via serial interface (optional)

Applications

  • aerosol generation from pure liquids, solutions and suspensions
  • integrity and recovery testing of HEPA/ULPA filters according to ISO 14644-3
  • acceptance measurements for clean rooms, safety work benches and smoke detectors
  • generation of calibration aerosols (e.g.: for particle counters) at low air flow rates

Accessories

  • battery connection cable / interface cable
  • particle instruments interface modul (terminal program)
  • aerosol substance (DEHS, PAO, PSL)
  • diffusion dryer (DDU 570/L, DDU 570/H)
  • aerosol dilution system (DIL 554), stainless steel, battery operation
  • isokinetic rectangular sampling probe (SYS 529) according to EN ISO 14644-3 and VDI 2083-3
  • cleanroom particle counter (LAP 340)
  • software for cleanroom validation (CRQWin)
  • carry case
Parameter titleUnitValue
setting parameter hPa differential nozzle pressure
setting range hPa 0 ... 1400
setting resolution hPa 1 (≤ 50), 5 (> 50)
setting method - manual, controlled
aerosol, volumetric flow rate L/h 20 ... 250
aerosol, mass flow rate (aerosol substance) g/h 0 ... 1,4
operation medium, gas/air - ambient air (filtered internally)
operation medium, aerosol substance - liquids (DEHS, PAO,...), solutions (NaCl, KCl,...), suspensions (PSL,...)
operating medium, aerosol substance - filling volume mL max. 80
power supply - 100 ... 240 V AC, 12 V DC (optional battery operation)
hose connector, outlet mm Ø 8
power consumption W max. 36
environmental condition, max. counterpressure kPa 20
dimensions (w × h × d) mm 120 × 195 × 300
weight kg 3,9