Select your language

Climomaster Anemometer 6501 Series

Manufacturer: Kanomax
User Manual
Data Sheet
Probe Engineering Drawings
Application Note - Anemometer Selection Guide
Application Note - Evolution of Hotwire Probes
Application Note - The Relationship Between Air Velociy, Temperature and Pressure
Application Note - Measuring Airflow in the Cleanroom
Software - Manual
Application Note - Anemometer Selection Guide
Application Note - Evolution of Hotwire Probes
Application Note - The Relationship Between Air Velociy, Temperature and Pressure
Application Note - Measuring Airflow in the Cleanroom
Software - Manual

Precision Engineered for Unmatched Accuracy

The Kanomax Climomaster 6501 Series stands as a testament to precision engineering, offering an unparalleled 2% m.v. accuracy level. This exceptional accuracy earmarks it as the world’s most precise hot-wire airflow meter, ideal for research-grade applications.

Versatile and Interchangeable Probe Options

This anemometer boasts remarkable flexibility with a suite of eight different probe types, each meticulously designed for specific environments. Whether it’s measuring temperature, humidity, turbulent or high-velocity airflow, or in confined spaces like circuit board cooling profiles, these probes are fully interchangeable with a single main unit, requiring no additional calibration, thus ensuring seamless adaptability.

Robust Data Management

With a substantial storage capacity for 20,000 records and advanced data logging capabilities, the Climomaster 6501 ensures detailed and extensive data collection. For enhanced functionality, optional features like a pressure sensor and analog output are available upon request.

Designed for Various Applications

The Climomaster 6501 is the go-to instrument for a range of critical applications. Its diverse utility spans HVAC Testing, Indoor Air Quality (IAQ) Investigations, Critical Environment Certification, HEPA Filter Performance Testing, Turbulent Air Testing, and Manufacturing Process Controls.

Comprehensive Accessory Range

To bolster its performance, the Climomaster 6501 Series comes with an array of additional accessories. These include software for data analysis, extension rods for hard-to-reach areas, a hands-free case for convenience, and specialized cable lengths to suit varied requirements.

Automatic Barometric Pressure Compensation

A standout feature of the Climomaster 6501 is its automatic barometric pressure compensation. This ensures that the device maintains its high accuracy even in extreme altitudes, making it a reliable tool in diverse environmental conditions.

Features and Benefits

  • Its sleek, ergonomic design with anti-slip grip comes standard with a highly visible backlit display making it the handheld choice of Journeymen technicians in the field.
  • Stores up to 20,000 data points which is easily downloaded to your company computer for analysis via the USB interface.
  • The optional software allows the users to view real-time data graphically formatted on their company computer
  • For your convenience the anemometer keeps track of its own calibration cycle and reminds you when it is due for calibration*.
  • The Climomaster 6501-CE measures pressure and features an analog output.
  • Equipped with an automatic atmospheric pressure compensation function.
  • When the direction of the airflow is known, choose a unidirectional probe.
  • When the direction of the airflow is unknown or turbulent, choose an omnidirectional probe.

*Annual NIST TRACEABLE CALIBRATIONS are performed at our facility in Andover, NJ.

Probes

Probe Model6531-2G6541-2G6561-2G6542-2G6533-2G6543-2G6551-2G6552-2G
Air Velocity Range (FPM) 2 to 6000 2 to 6000 2 to 9840 2 to 6000 2 to 1000 2 to 1000 2 to 6000 2 to 6000
Airflow Direction Uni Uni Uni Omni Omni Omni Omni Omni
Probe Type Rod Rod Rod Needle Spherical Spherical Miniature Miniature
Air Velocity O O O O O O O O
Temperature O O O O O O
Relative Humidity O O

Accessories

  • Uni-directional Rod Probe (V/T/H): Model 6531-2G
  • Omni-directional Spherical Probe (V/T/H): Model 6533-2G
  • Uni-directional Rod Probe (V/T): Model 6541-2G
  • Omni-directional Needle Probe (V/T): Model 6542-2G
  • Omni-directional Spherical Probe (V/T): Model 6543-2G
  • Omni-directional Mini-Spherical Probe (V): Model 6551-2G
  • Omni-directional Mini-Spherical Probe with 90° Radius Bend (V): Model 6552-2G
  • Uni-directional Rod Probe (V/T): Model 6561-2G
  • Probe Flexible Extension Rod: Model 6531-04
  • Probe Telescopic Extension Rod: Model 6531-05
  • Portable Printer: Model DPU-S245
  • Portable Printer Paper: Model TP-5RLPK
  • 6501 Series Hands-Free Case: Model 6000-63
  • AC Adapter: Model 6113-02
  • PC Software: Model 6000-41
  • 2M Probe Cable: Model 6531-06
  • 5M Probe Cable: Model 6531-09
  • 10M Probe Cable: Model 6531-10
  • 20M Probe Cable: Model 6531-11

Climomaster Main Unit Specifications

Main Unit Specifications
Differential Pressure
Range -5 to 5 kPa
Resolution 0.01 kPa
Accuracy ±3% of READING or ±0.01 kPa whichever is greater
General Specifications
Analog Output DC 0 to 1 V
Interface USB / Printer
Datalogging Stores up to 20000 measurement data
Power Supply 6 × AA Batteries or AC Adapter (AC 100 to 240 V , 50/60 Hz)
Weight 0.9 lbs (400 g)
Operating Environment 41 to 104°F (5 to 40°C) with no condensation
Storage Environment 14 to 122°F (-10 to 50°C) with no condensation
Standards CE-mark
Warranty 2 year
What’s Included Meter, Operation Manual, Carrying Case, USB Cable, AA Batteries, 2m Probe Cable

 

Unit Model6501-0E6501-CE
Differential Pressure O
Analog Output O
USB Interface O O
Printer Interface O O

Climomaster Probe Specifications

Measurement Specifications
Air Velocity
Range 2 to 9840 FPM (0.01 to 50 m/s)
Resolution 1 FPM (0.01 m/s)
Accuracy +/- 2% of READING or +/- 0.015 m/s (+/- 3 FPM) whichever is greater
Temperature
Range -4 to 158°F (-20 to 70°C)
Resolution 1°F (0.1°C)
Accuracy +/- 1°F (+/- 0.5 °C)
Relative Humidity
Range 2.0 to 98.0 %RH
Resolution 0.1%
Accuracy ±2.0% (at range of 2 to 80%) ±3.0% (at range of 80 to 98%)
General Specifications
Operating Environment -4 to 158°F (-20 to 70°C) with no condensation
Storage Environment 14 to 122°F (-10 to 50°C) with no condensation
Standards CE-mark
Warranty 2 year
Standard Accessories Probe, NIST Certificate

 

Probe Model6531-2G6541-2G6561-2G6542-2G6533-2G6543-2G6551-2G6552-2G
Air Velocity Range (FPM) 2 to 6000 2 to 6000 2 to 9840 2 to 6000 2 to 1000 2 to 1000 2 to 6000 2 to 6000
Airflow Direction Uni Uni Uni Omni Omni Omni Omni Omni
Probe Type Rod Rod Rod Needle Spherical Spherical Miniature Miniature
Air Velocity O O O O O O O O
Temperature O O O O O O
Relative Humidity O O
Copyright MAXXmarketing GmbH