Dilution System DIL 546/C

Manufacturer: Topas
product sheet DIL 500 series

The self-adjusting DIL 546/C Dilution System is used to defined and reproducible adjustment of the aerosol particle concentration (dilution).

Standards

VDI 3491-6
ISO 14644-3

It is particularly useful for the verification of the test aerosol for testing cleanroom installations. In the standard EN ISO 14644-3, this is included as a requirement.

Benefits

  • dilution ratio is automatically adjusted
  • mobile use (no compressed air supply is required, battery operation possible)
  • directing the entire intake volume flow of the particle counter from the sampling point via the dilution system (closed system)
  • constant, reproducible dilution even under changing operating conditions
  • internal monitoring of the dilution ratio with display
  • high durability and reliability, minimum maintenance
  • over- and under-pressure operation is permitted
  • adaptation to customer-specific requirements (dilution ratios) possible
  • can be operated in connection of series

Applications

  • verification of the test aerosol for testing cleanroom installations
  • acceptance measurements of laminar flow boxes, safety cabinets and smaller clean rooms
  • determination of filter separation efficiencies
  • measurement of highly concentrated aerosols
  • aerosol research
Parameter titleUnitValue
purpose of conditioning - dilution of aerosols
sample condition, outlet - diluted aerosol
dilution factor - 10
aerosol, volumetric flow rate L/min 2,83 (3,0)
setting method - automatic, controlled
sample condition, inlet - non-diluted aerosol
power supply - 12 V DC; 0,1 A (optional battery operation)
hose connector, outlet mm Ø 8 (optional Ø 10)
pressure drop kPa 0,8 ... 1,3
environmental condition, temperature °C 10 ... 50
environmental condition, max. counterpressure kPa ± 20
dimensions (w × h × d) mm 110 × 160 × 300
weight kg 1,5